    "data_type": "CVE",
    "data_format": "MITRE",
    "data_version": "4.0",
    "CVE_data_meta": {
        "ID": "CVE-2020-12045",
        "ASSIGNER": "ics-cert@hq.dhs.gov",
        "STATE": "PUBLIC"
    },
    "affects": {
        "vendor": {
            "vendor_data": [
                {
                    "vendor_name": "n/a",
                    "product": {
                        "product_data": [
                            {
                                "product_name": "Baxter Sigma Spectrum Infusion Pumps",
                                "version": {
                                    "version_data": [
                                        {
                                            "version_value": "Sigma Spectrum v6.x model 35700BAX,Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module (WBM) v9,v11,v13,v14,v15,v16,v20D29,v20D30,v20D31,v22D24, Spectrum v8.x w/WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum LVP v8.x with WBM v17, v20D29,v20D30,v20D31,and v22D24"
                                        }
                                    ]
                                }
                            }
                        ]
                    }
                }
            ]
        }
    },
    "problemtype": {
        "problemtype_data": [
            {
                "description": [
                    {
                        "lang": "eng",
                        "value": "USE OF HARD CODED PASSWORD CWE-259"
                    }
                ]
            }
        ]
    },
    "references": {
        "reference_data": [
            {
                "refsource": "MISC",
                "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-04",
                "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-04"
            }
        ]
    },
    "description": {
        "description_data": [
            {
                "lang": "eng",
                "value": "The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials."
            }
        ]
    }
}