    "CVE_data_meta": {
        "ASSIGNER": "cve@mitre.org",
        "ID": "CVE-2019-11687",
        "STATE": "PUBLIC"
    },
    "affects": {
        "vendor": {
            "vendor_data": [
                {
                    "product": {
                        "product_data": [
                            {
                                "product_name": "n/a",
                                "version": {
                                    "version_data": [
                                        {
                                            "version_value": "n/a"
                                        }
                                    ]
                                }
                            }
                        ]
                    },
                    "vendor_name": "n/a"
                }
            ]
        }
    },
    "data_format": "MITRE",
    "data_type": "CVE",
    "data_version": "4.0",
    "description": {
        "description_data": [
            {
                "lang": "eng",
                "value": "An issue was discovered in the DICOM Part 10 File Format in the NEMA DICOM Standard 1995 through 2019b. The preamble of a DICOM file that complies with this specification can contain the header for an executable file, such as Portable Executable (PE) malware. This space is left unspecified so that dual-purpose files can be created. (For example, dual-purpose TIFF/DICOM files are used in digital whole slide imaging for applications in medicine.) To exploit this vulnerability, someone must execute a maliciously crafted file that is encoded in the DICOM Part 10 File Format. PE/DICOM files are executable even with the .dcm file extension. Anti-malware configurations at healthcare facilities often ignore medical imagery. Also, anti-malware tools and business processes could violate regulatory frameworks (such as HIPAA) when processing suspicious DICOM files."
            }
        ]
    },
    "problemtype": {
        "problemtype_data": [
            {
                "description": [
                    {
                        "lang": "eng",
                        "value": "n/a"
                    }
                ]
            }
        ]
    },
    "references": {
        "reference_data": [
            {
                "url": "https://labs.cylera.com/2019.04.16/pe-dicom-medical-malware",
                "refsource": "MISC",
                "name": "https://labs.cylera.com/2019.04.16/pe-dicom-medical-malware"
            },
            {
                "refsource": "MISC",
                "name": "https://github.com/d00rt/pedicom/blob/master/doc/Attacking_Digital_Imaging_and_Communication_in_Medicine_(DICOM)_file_format_standard_-_Markel_Picado_Ortiz_(d00rt).pdf",
                "url": "https://github.com/d00rt/pedicom/blob/master/doc/Attacking_Digital_Imaging_and_Communication_in_Medicine_(DICOM)_file_format_standard_-_Markel_Picado_Ortiz_(d00rt).pdf"
            },
            {
                "refsource": "MISC",
                "name": "https://github.com/d00rt/pedicom",
                "url": "https://github.com/d00rt/pedicom"
            },
            {
                "refsource": "BID",
                "name": "108730",
                "url": "http://www.securityfocus.com/bid/108730"
            }
        ]
    }
}